hazmat meters

  1. Image Upload 2
    HazMat CAD Plus
  2. what are the two majpr categories of agents that the Hazmat CAD plus reads for?
    • Toxic Industrail Chemicals (TICS)
    • Chemical warfare agents (CW)
  3. What two modes does the Hazmat CAD Plus operate in and what are the analysis times for each?
    • Fast Mode (FM) 20 second analysis time
    • High Sensitivity (HS) 120 second analysis time
  4. -Hazmat CAD Plus-
    what are the 3 positions and their purposes on the sample inlet?
    • Open - for stimulant testing
    • Closed - for storage
    • Dust Filter - impedes particulates and water from entering inlet. this position should be used for field sampling
  5. -Hazmat CAD Plus-
    The LED status display area shows; Alarm, All-Clear, LoBat, Sys OK, what do each of these mean?
    • Alarm - red blinking light indicates a threat vapor has been detected
    • All Clear - green steady light no vapor threat present
    • Lo Bat - yellow light indicates power is running low
    • Sys OK - Green blinking light indicates proper electrical operation of instrument
  6. -Hazmat CAD Plus-
    What type of battery does the instrument take?
    Sony Lithium Ion battery (sony video camera battery)
  7. -Hazmat CAD Plus-
    What are "G" agents
    "G" are nerve agents

    • GA - Tabun
    • GB - Sarin
    • GD - Soman
    • GF - No common name
  8. -Hazmat CAD Plus-
    What are "H" agents?
    • "H" are blister agents
    • Mustard gases - sulfer mustard or nitrogen mustard
  9. -Hazmat CAD Plus-
    Can the meter detect VX (nerve agent, same category as "G")
    No Not pure VX. It will detect technical grade (impure) so it is not a complete solution for detecting VX.
  10. -Hazmat CAD Plus-
    What are the 7 chemical agent codes that could be displayed on the meter?
    • G
    • H
    • TOX
    • BLoD - Hydrogen Cyanide
    • CHOK - Phosgene Gas
    • HYdR- Hydride Gas
    • HALO- Halogen Gas
  11. -Hazmat CAD Plus-
    When the detector identifies a chemical agent the alarm will give on of three concentration levels. What are they and what are the values?
    • Low - the detection devices (saw) preset alarm threshold
    • Medium - 2 times higher than the preset alarm threshold (low)
    • High - 5 times higher than the preset threshold (low)
  12. -Hazmat CAD Plus-
    High Sensitivity Mode will detect concentration levels ____ lower than Fast Mode?
    6 times
  13. -Hazmat CAD Plus-
    How do you perform a confidence check and when should this be done?
    - Meter must be running for 15 min and used in FAST MODE with all appropriate green lights illuminated.  Use diffusion tubes by placing in sample inlet for 3-5 seconds.  a "G" alarm should occur.  This provides a TOTAL system performance check.

    - Chlorine Bleach on a cotton swab can be used to check a response for Halogen exposure

    - This should be done once a month and prior to any entry
  14. - GAS RANGER - 
    What mode would you use to pinpoint the source of a gas leak?
    Sniff/Track Gas
  15. - GAS RANGER - 
    5% gas by volume - How many PPM?
    • 50,000 PPM
    • 10,000 per each 1%
  16. - GAS RANGER - 
    The meter alarms at what % LEL and what % GAS?
    • LEL 10%
    • GAS 5%
  17. - GAS RANGER - 
    What will the display indicate if the battery is low?
    Flashes "LOW"
  18. Image Upload 4
    Toxirae Pro
  19. - TOXI RAE METER -
    What do we have to know prior to using this meter?
    What gas we are dealing with. The meter is a gas specific meter
  20. - TOXI RAE METER -
    What three gas sensors are supplied for this meter?
    Ammonia NH3, Chlorine Cl2, Cyanide HCN
  21. - TOXI RAE METER -
    What powers the meter?
    a lithium Ion rechargeable battery.  The Toxi Rae should be kept on the charger when not in use.
  22. - TOXI RAE METER -
    What can you do if the meter fails to stay powered up?
    Battery is dead .. Recharge the unit as soon as possible
  23. - TOXI RAE METER -
    The meter is passive.  What does this mean?
    The meter does not have a pump to pull the gas through the meter. Therefore, you cannot connect a hose or a wand to the meter.
  24. - TOXI RAE METER -
    What are the alarm indication codes?
    • MAX - Gas exceeds electronics maximum range
    • OVER - Gas exceeds measurement range
    • HIGH - Gas exceeds high alarm limit
    • LOW - Gas exceeds low alarm limit
    • STEL - Gas exceeds Short term exposure limit
    • TWA - Gas exceeds time weighted average limit
    • CAL - calibration failed or needs calibration
    • BAT - low battery
  25. - TOXI RAE METER -
    How can you find the measurement ranges for each sensor?
    The meter will tell you the measurement ranges for each sensor during the warm up cycle
  26. - TOXI RAE METER -
    What are some key points concerning the Smart Cells?
    • -  Cells last about 2 years
    • -  Exposure to high temps or moisture can be harmful
    • - Exposure to high concentrations can damage cell
    • - Always keep sensors in keeper, in the meter, or on the charger. They can die if they go too long without electrical charge
    • - cells exposed to very high concentrations of gas may take hours to clear or can be permanently damaged.  Must be left running until cleared
  27. - TOXI RAE METER -
    Explain how to change sensors.
    The manufacturer recommendation is to turn off the meter to change out the sensor.
  28. - TOXI RAE METER -

    Name 5 sensors that can be used by this meter
    • Ammoinia,
    • Chlorine,
    • Hydrogen Cyanide,
    • Nitrogen Dioxcide,
    • Sulfur Dioxcide
  29. - TOXI RAE METER -

    Can you change the sensor without turning off the meter?
    No (Manufacturers Recommendations)
  30. - TOXI RAE METER -

    1000 ppm equals what percent by volume of air?
    .1%
  31. - TOXI RAE METER -

    What kind of battery does the Toxi RAE use?
    Lithium Rechargeable
  32. - TOXI RAE METER -

    Can a Carbon Dioxide sensor be used in a Toxi RAE meter?
    No, it can only be used in a CO2 meter.
  33. Image Upload 6
    Multi Rae (5 gas)
  34. - MULTI RAE (5 GAS) -

    The 5 areas of atmospheric measurement are?
    • Combustibility
    • Oxygen O2
    • Carbon Monoxide CO
    • Hydrogen Sulfide H2S
    • VOCs
  35. - MULTI RAE (5 GAS) - 

    Can other sensors other than the standard 5 be used in a Multi RAE meter?
    yes, multiple sensors can be used
  36. - MULTI RAE (5 GAS) - 

    What are the action / Alarm levels?
    • LEL alarms at 10%
    • O2 alarms at 19.5 and 23.5%
    • CO alarms at 35 PPM
    • H2S alarms at 10 PPM
    • VOC alarms at 50-100 PPM
  37. - MULTI RAE (5 GAS) - 

    20% by volume - how many PPM``
    • 200,00 PPM 
    • 10,000 per 1%
  38. - MULTI RAE (5 GAS) - 

    What type of battery does the Multi RAE pro use?
    Rechargeable Lithium Ion Battery.  Meter should be kept on the charger when not in use.
  39. - MULTI RAE (5 GAS) - 

    When air monitoring around electrical hazards what type of sampling wand is used?`
    Non Conductive Wand
  40. Image Upload 8
    Jerome Meter
  41. - JEROME METER - 

    The measurement of when an area or person is considered contaminated?
    .03 mg/mm3
  42. - JEROME METER - 

    What do the displays mean?
    LOBAT
    EEE 
    PPP
    LLL
    888 Meter
    HL
    • - LOBAT Battery needs to be recharged
    • - EEE Battery to low to function (turms off)
    • - PPP Power Cord Required
    • - LLL Meter needs to be re-zero
    • - 888 Meter needs to be regenerated
    • - HL Sample detected exceeds maximum ranges of meter .999 mg/m3 Hg
  43. - JEROME METER - 

    Identify what the Jerome meter test/analyzes
    Mercury vapors in the low mg/mm3