what are the two majpr categories of agents that the Hazmat CAD plus reads for?
- Toxic Industrail Chemicals (TICS)
- Chemical warfare agents (CW)
-
What two modes does the Hazmat CAD Plus operate in and what are the analysis times for each?
- Fast Mode (FM) 20 second analysis time
- High Sensitivity (HS) 120 second analysis time
-
-Hazmat CAD Plus-
what are the 3 positions and their purposes on the sample inlet?
- Open - for stimulant testing
- Closed - for storage
- Dust Filter - impedes particulates and water from entering inlet. this position should be used for field sampling
-
-Hazmat CAD Plus-
The LED status display area shows; Alarm, All-Clear, LoBat, Sys OK, what do each of these mean?
- Alarm - red blinking light indicates a threat vapor has been detected
- All Clear - green steady light no vapor threat present
- Lo Bat - yellow light indicates power is running low
- Sys OK - Green blinking light indicates proper electrical operation of instrument
-
-Hazmat CAD Plus-
What type of battery does the instrument take?
Sony Lithium Ion battery (sony video camera battery)
-
-Hazmat CAD Plus-
What are "G" agents
"G" are nerve agents
- GA - Tabun
- GB - Sarin
- GD - Soman
- GF - No common name
-
-Hazmat CAD Plus-
What are "H" agents?
- "H" are blister agents
- Mustard gases - sulfer mustard or nitrogen mustard
-
-Hazmat CAD Plus-
Can the meter detect VX (nerve agent, same category as "G")
No Not pure VX. It will detect technical grade (impure) so it is not a complete solution for detecting VX.
-
-Hazmat CAD Plus-
What are the 7 chemical agent codes that could be displayed on the meter?
- G
- H
- TOX
- BLoD - Hydrogen Cyanide
- CHOK - Phosgene Gas
- HYdR- Hydride Gas
- HALO- Halogen Gas
-
-Hazmat CAD Plus-
When the detector identifies a chemical agent the alarm will give on of three concentration levels. What are they and what are the values?
- Low - the detection devices (saw) preset alarm threshold
- Medium - 2 times higher than the preset alarm threshold (low)
- High - 5 times higher than the preset threshold (low)
-
-Hazmat CAD Plus-
High Sensitivity Mode will detect concentration levels ____ lower than Fast Mode?
6 times
-
-Hazmat CAD Plus-
How do you perform a confidence check and when should this be done?
- Meter must be running for 15 min and used in FAST MODE with all appropriate green lights illuminated. Use diffusion tubes by placing in sample inlet for 3-5 seconds. a "G" alarm should occur. This provides a TOTAL system performance check.
- Chlorine Bleach on a cotton swab can be used to check a response for Halogen exposure
- This should be done once a month and prior to any entry
-
- GAS RANGER -
What mode would you use to pinpoint the source of a gas leak?
Sniff/Track Gas
-
- GAS RANGER -
5% gas by volume - How many PPM?
- 50,000 PPM
- 10,000 per each 1%
-
- GAS RANGER -
The meter alarms at what % LEL and what % GAS?
-
- GAS RANGER -
What will the display indicate if the battery is low?
Flashes "LOW"
-
-
- TOXI RAE METER -
What do we have to know prior to using this meter?
What gas we are dealing with. The meter is a gas specific meter
-
- TOXI RAE METER -
What three gas sensors are supplied for this meter?
Ammonia NH3, Chlorine Cl2, Cyanide HCN
-
- TOXI RAE METER -
What powers the meter?
a lithium Ion rechargeable battery. The Toxi Rae should be kept on the charger when not in use.
-
- TOXI RAE METER -
What can you do if the meter fails to stay powered up?
Battery is dead .. Recharge the unit as soon as possible
-
- TOXI RAE METER -
The meter is passive. What does this mean?
The meter does not have a pump to pull the gas through the meter. Therefore, you cannot connect a hose or a wand to the meter.
-
- TOXI RAE METER -
What are the alarm indication codes?
- MAX - Gas exceeds electronics maximum range
- OVER - Gas exceeds measurement range
- HIGH - Gas exceeds high alarm limit
- LOW - Gas exceeds low alarm limit
- STEL - Gas exceeds Short term exposure limit
- TWA - Gas exceeds time weighted average limit
- CAL - calibration failed or needs calibration
- BAT - low battery
-
- TOXI RAE METER -
How can you find the measurement ranges for each sensor?
The meter will tell you the measurement ranges for each sensor during the warm up cycle
-
- TOXI RAE METER -
What are some key points concerning the Smart Cells?
- - Cells last about 2 years
- - Exposure to high temps or moisture can be harmful
- - Exposure to high concentrations can damage cell
- - Always keep sensors in keeper, in the meter, or on the charger. They can die if they go too long without electrical charge
- - cells exposed to very high concentrations of gas may take hours to clear or can be permanently damaged. Must be left running until cleared
-
- TOXI RAE METER -
Explain how to change sensors.
The manufacturer recommendation is to turn off the meter to change out the sensor.
-
- TOXI RAE METER -
Name 5 sensors that can be used by this meter
- Ammoinia,
- Chlorine,
- Hydrogen Cyanide,
- Nitrogen Dioxcide,
- Sulfur Dioxcide
-
- TOXI RAE METER -
Can you change the sensor without turning off the meter?
No (Manufacturers Recommendations)
-
- TOXI RAE METER -
1000 ppm equals what percent by volume of air?
.1%
-
- TOXI RAE METER -
What kind of battery does the Toxi RAE use?
Lithium Rechargeable
-
- TOXI RAE METER -
Can a Carbon Dioxide sensor be used in a Toxi RAE meter?
No, it can only be used in a CO2 meter.
-
-
- MULTI RAE (5 GAS) -
The 5 areas of atmospheric measurement are?
- Combustibility
- Oxygen O2
- Carbon Monoxide CO
- Hydrogen Sulfide H2S
- VOCs
-
- MULTI RAE (5 GAS) -
Can other sensors other than the standard 5 be used in a Multi RAE meter?
yes, multiple sensors can be used
-
- MULTI RAE (5 GAS) -
What are the action / Alarm levels?
- LEL alarms at 10%
- O2 alarms at 19.5 and 23.5%
- CO alarms at 35 PPM
- H2S alarms at 10 PPM
- VOC alarms at 50-100 PPM
-
- MULTI RAE (5 GAS) -
20% by volume - how many PPM``
-
- MULTI RAE (5 GAS) -
What type of battery does the Multi RAE pro use?
Rechargeable Lithium Ion Battery. Meter should be kept on the charger when not in use.
-
- MULTI RAE (5 GAS) -
When air monitoring around electrical hazards what type of sampling wand is used?`
Non Conductive Wand
-
-
- JEROME METER -
The measurement of when an area or person is considered contaminated?
.03 mg/mm3
-
- JEROME METER -
What do the displays mean?
LOBAT
EEE
PPP
LLL
888 Meter
HL
- - LOBAT Battery needs to be recharged
- - EEE Battery to low to function (turms off)
- - PPP Power Cord Required
- - LLL Meter needs to be re-zero
- - 888 Meter needs to be regenerated
- - HL Sample detected exceeds maximum ranges of meter .999 mg/m3 Hg
-
- JEROME METER -
Identify what the Jerome meter test/analyzes
Mercury vapors in the low mg/mm3
milligram per cubic millimeter
|
|