APMDA boutelle L3&L4

  1. How does microdialysis work, with a diagram?
    • Image Upload 2
    • 1. Solution pumped into the probe is sterile water called perfusate
    • 2. It washes over the inside of the semi-permeable membrane tip of the microdialysis probe
    • 3. Chemical speceies in the extracellular fluid that are not in the perfusate  in the same concentration, DIFFUSE down the concentration gradient, cross the semi-permeable membrane and are then carried away form the membrane area to give the dialysate.
    • 4. Dialysate collected for ex-vivo analysis
  2. Factors affecting probe recovery in vitro
    If solution is well stirred, concentration gradient appears across the membrane

    Dialysate concentration limited by diffusive flux and internal flow rate.

    In vitro, you need a vortex created by high speed stirring. The conc grad is across the memb only.
  3. maybe delete this one...Factors affecting microdialysis recovery
    • Passive effects:
    • Excluded volume and totuous difffusion path
    • Tissue effects are rate limiting
    • Prove recovery<invitro

    • Active effects:
    • Release rate to local ECF conc
    • Buffering of extracellular environment
    • In vitro>probe recovery> in vivo passive
  4. How is passive in vivo recovery compared to in vitro case and why
    • Results in Probe recovery < in vitro case
    • Image Upload 4
    • Excluded volume and totuous difffusion path
    • Lack of stirring
    • Tissue effects are rate limiting


    These molecules therefore diffuse to the probe much less efficiently. Hence the concentration is lower and probe recovery is lower than in vitro
  5. What is the probe recovery of the in-vivo active case like compared to in vivo passive and in vitro? Why?
    • In vitro> probe recovery in vivo active> in vivo passive
    • Image Upload 6
    • This is due to the tissue actively responds to a drop in local concentration by releasing more and uptaking less. Summarised as

    - Buffering of extracellular environment


    This causes the diffusion projle to move towards the in vitro value
  6. State what two methods are there for measuring true  tissue levels
    • Zero nett flux
    • Extrapolation to zero flow.
  7. How does zero nett flux work?
    • Graphically find the point at which the microdialysis probe neither removes or adds molecules to the ECF. 
    • The slope of the graph is the probe recovery E
    • The y intercept is the normal mode microdialysis conc.
    • The x-intercept is the ECF conc
    • Vary Cin and measure Cout
  8. How does extrapolation to zero flow work?
    • The flow rate through probe is reduced and dialysate conc measured
    • Process continued until dialysate conc = true ECF value

    (so youre like plotting this decaying exponential until you get to 100% which takes ages)
  9. What is the problem with zero nett flux?
    It can have errors if there is significant damaged tissue
  10. How does one do microdialysis?
    • Insert probes slowly (0.3-4uL/s): to minimise tissue damage
    • Allow 12 hrs for recovery: 
    • Perfuse with correct ionic solution: tissue is sensitive to ionic composition of perfusate. If it is incorrect, the tissue might be more/less electrically active. 
    • Check signal integrity
  11. How does one check signal integrity when doing microdialysis
    Use TTX to lock voltage gated sodium channels.
  12. How does one perform good experimental design with a microdialysis experiment? What statistics approach does this let us use?
    • 1. Monitor stimulated response and ensure it is reproduceable
    • 2. Challenge with local drug through probe
    • 3. Check that response returns on drug washout. 

    This lets us use a pairwise statistics approach: Lets us compare control and drug response in the same animal.
  13. How does one do high performance liquid chromatography combined with Electochemical detection?
    • 1. Collect a sample: Takes 10-20 mins. Stored in an autosampler.
    • 2. Use chromatography to separate the different components
    • 3. analyse the peaks as they elute from the column
    • 4. If the levels are too low inject more sample- hence its mass sensitive.
  14. What are the potential problems with microdialysis?
    • The probe is much larger than cells
    • The number of intact neurons can be reduced which is important if you want to measure substances with short lifetimes and hence short diffusional lengths.

    • Second example
    • The probe caused a proliferation of glial cells to occur close to the membrane.
    • The consequence of this is that you get a lower proportion of neurons
  15. How does facilitated recovery make microdialysis better?
    • Idea is to make capture of released molecules more effective by trapping them on antibody-coated nano-particles
    • They bind the molecule targeted by the antibody very tightly.
    • This leaves a very low 'free' concentration left in the dialysate
    • Additionaly, it stops 'sticky' molecules such as peptides from sticking to the connecting tubing in non-specific binding.
  16. Explain the problem with increasing sampling rate that occurs with microdialysis. Whats the solution?
    As temporal resolutio increases, moles available to detect decrease, but the number of samples dramatically increases.

    You have to use an automatic sampling to use high temporal resolution
  17. What are the advantages of in-vivo electrodes
    • Very fast response ( much less dead volume to allow faster changes to be seen)
    • Very sensitive
    • Tissue damage is small
  18. How does the carbon fibre in vivo electrode work?
    • Fast voltage ramp applied to electrode over 10ms 
    • Dopamine oxided to dopamine-o-puinone to give a current peak
    • On the reverse scan, the dopamine-o-quinone is reduced back to dopamine to give a reverse peak.
    • Perform background scan which is subsequently removed from subsequent scans
    • Position of oxidation reduction peak is diagnostic for the chemical species
    • The positions of the peaks is diagnostic for the anlyste.
  19. What the disadvantages of carbon fibre electrodes
    • Can only detect species that can be oxidised or reduced by the electrode
    • To get a signal, you must subtract the baseline signal, so you cannot measure basal concentrations.
  20. What are the limiting factors in how frequently you can analyse a sample?
    • 1. Ability to handle small volumes
    • 2. Ability to handle large number of samples
    • 3. Analytical sensitivity of the detection system.
Author
keesukim
ID
339527
Card Set
APMDA boutelle L3&L4
Description
microdialysis stuff
Updated