Focal-spot blur is greatest
A
toward the anode end of the x-ray beam
B
toward the cathode end of the x-ray beam
C
directly along the course of the CR
D
as the SID is increased
 -
- B
-
- toward the cathode end of the x-ray beam
-
Which of the following will occur as a result of a decrease in the anode target angle?
1. |
Less pronounced anode heel effect |
2. |
Decreased effective focal spot size |
3. |
Greater photon intensity toward the cathode side of the x-ray tube |
A
1 only
B
1 and 2 only
C
2 and 3 only
D
1, 2, and 3
 -
- C
-
- 2 and 3 only
-
The output phosphor can be coupled with the Vidicon TV camera or charge-coupled device (CCD) via
- fiber optics.
- an image distributor or lens.
- closed-circuit TV.
A
1 only
B
1 and 2 only
C
2 and 3 only
D
1, 2, and 3
 -
- B
-
- 1 and 2 only
-
f the distance from the focal spot to the center of the collimator's mirror is 6 in., what distance should the illuminator's light bulb be from the center of the mirror?
A
3 in.
B
6 in.
C
9 in.
D
12 in.
 -
- B
-
- 6 in.
-
Off-focus, or extrafocal, radiation is minimized by
A
avoiding the use of very high kilovoltages
B
restricting the x-ray beam as close to its source as possible
C
using compression devices to reduce tissue thickness
D
avoiding extreme collimation
 -
- B
-
- restricting the x-ray beam as close to its source as possible
-
Although the stated focal-spot size is measured directly under the actual focal spot, focal-spot size in fact varies along the length of the x-ray beam. At which portion of the x-ray beam is the effective focal spot the largest?
A
At its outer edge
B
Along the path of the CR
C
At the cathode end
D
At the anode end
 -
- C
-
- At the cathode end
-
What lies immediately under the phosphor layer of a PSP storage plate?
A
Reflective layer
B
Base
C
Antistatic layer
D
Lead foil
 -
- A
-
- Reflective layer
-
Just under the barium fluorohalide layer is a____that helps direct emitted light up toward the CR reader.
reflective layer
-
Below the reflective layer is the ____, behind that is an ____ layer, and then the lead foilto absorb backscatter.
base
antistatic
-
Over the top of the barium fluorohalide is a ____ layer.
protective
-
The batteries in battery-operated mobile x-ray units provide power to
- the x-ray tube
- machine locomotion
- the braking mechanism
A
1 only
B
2 only
C
1 and 2 only
D
1, 2, and 3
 -
- D
-
- 1, 2, and 3
-
Image quality in digital fluoroscopy is influenced by
1. |
pixel size. |
2. |
contrast. |
3. |
noise. |
A
1 only
B
1 and 2 only
C
2 and 3 only
D
1, 2, and 3
 -
- D
-
- 1, 2, and 3
-
_______ is degrading to both traditional and digital images. It can result from a number of causes including insufficient mA
Noise
-
Which of the following voltage ripples is (are) produced by single-phase equipment
1. |
100% voltage ripple |
2. |
13% voltage ripple |
3. |
3.5% voltage ripple |
A
1 only
B
2 only
C
2 and 3 only
D
1, 2, and 3
 -
- A
-
- 1 only
-
Three-phase, 6-pulse equipment has a __% voltage ripple,
13
-
a three-phase, 12-pulse equipment has only a ___% ripple.
3.5
-
Three-phase, 12-pulse equipment comes closest to constant potential, as the voltage never falls below ___% of maximum value.
96.5
-
A focal-spot size of 0.3 mm or smaller is essential for which of the following procedures?
A
Bone radiography
B
Magnification radiography
C
Tomography
D
Fluoroscopy
 -
- B
-
- Magnification radiography
-
Which of the following combinations would pose the least hazard to a particular anode?
A
1.2-mm focal spot, 92 kVp, 1.5 mAs
B
0.6-mm focal spot, 80 kVp, 3 mAs
C
1.2-mm focal spot, 70 kVp, 6 mAs
D
0.6-mm focal spot, 60 kVp, 12 mAs
 -
- A
-
- 1.2-mm focal spot, 92 kVp, 1.5 mAs
-
Which of the following field sizes would require the least amount of patient X-ray dosage?
A
12 cm field
B
17 cm field
C
25 cm field
D
30 cm field
 -
- A
-
- 12 cm field
-
To maintain image clarity in an image-intensifier system, the path of electron flow from the photocathode to the output phosphor is controlled by
A
the accelerating anode
B
electrostatic lenses
C
the vacuum glass envelope
D
the input phosphor
 -
- B
-
- electrostatic lenses
-
Component parts of a CT imaging system include a(n)
- high-frequency generator
- x-ray tube
- operator console
A
1 only
B
1 and 2 only
C
1 and 3 only
D
1, 2, and 3
 -
- D
-
- 1, 2, and 3
-
Which of the following devices is used to control voltage by varying resistance?
A
Autotransformer
B
High-voltage transformer
C
Rheostat
D
Fuse
 -
- C
-
- Rheostat
-
The________ transformer increases the voltage and decreases the current.
high-voltage
-
The ____ is a type of variable resistor that is used to change voltage or current values.
rheostat
|
|