To determine how quickly an x-ray tube will disperse its accumulated heat, the radiographer uses a(n)
A
technique chart.
B
radiographic rating chart.
C
anode cooling curve.
D
spinning top test.
 -
- C
-
- anode cooling curve.
-
The ability of an x-ray unit to produce constant radiation output, at a given mA, using various combinations of mAs and time is called
A
linearity.
B
reproducibility.
C
densitometry.
D
sensitometry.
 -
- A
-
- linearity.
-
In digital imaging, artifacts arise from a number of sources, including which of the following?
A
Imaging hardware
B
Image processing software
C
Operator error artifacts
D
All of these may be sources of image artifacts
 -
- D
-
- All of these may be sources of image artifacts
-
All of the following are daily quality control (QC) steps to ensure optimal diagnostic images on a digital display monitor, except:
A
Turn on the monitor and allow it to warm up
B
Evaluate luminance, reflection, noise and glare
C
Make sure that the monitor is dust-free on the viewing surface
D
Evaluate the QC patterns at the periphery and verify that all letters and numbers appear
 -
- D
-
- Evaluate the QC patterns at the periphery and verify that all letters and numbers appear
-
The x-ray beam and collimator light field must coincide to within
A
10% of the OID
B
2% of the OID
C
10% of the SID
D
2% of the SID
 -
- D
-
- 2% of the SID
-
To compensate for variations in gain across a digital receptor, which of the following maintenance steps should be taken?
A
Conduct a calibration correction for image nonuniformity
B
Increase or decrease the exposure factors to compensate
C
Install a variable resistor to adjust the electrical supply to the unit
D
Keep a log for at least 30 days to confirm consistent variations before making any adjustments
 -
- A
-
- Conduct a calibration correction for image nonuniformity
-
A device used to measure the luminance response and uniformity of monitors used in digital imaging is called a
A
Penetrometer
B
Densitometer
C
Sensitometer
D
Photometer
 -
- D
-
- Photometer
-
Environmental background radiation can contribute to a
A
Ghost artifact
B
Moiré artifact
C
Static artifact
D
Grid cutoff artifact
 -
- A
-
- Ghost artifact
-
Too much edge enhancement of the image in digital systems can cause an unwanted artifact called the:
A
Photoelectric effect artifact
B
Scaling artifact
C
Hounsfield artifact
D
Halo effect artifact
 -
- D
-
- Halo effect artifact
-
All of the following are steps that should be used to accomplish quality control (QC) in digital radiography, except:
A
Acceptance testing
B
Establishment of baseline performance
C
Monitoring patient size to evaluate variations in equipment performance
D
Diagnosis of changes in performance
 -
- C
-
- Monitoring patient size to evaluate variations in equipment performance
-
An exposed image plate will retain its original image quality for about
A
2 hours
B
8 hours
C
24 hours
D
48 hours
 -
- B
-
- 8 hours
-
The device used to test the accuracy of the x-ray timer is the
A
densitometer
B
sensitometer
C
penetrometer
D
spinning top
 -
- D
-
- spinning top
-
To ensure proper operation of the digital image display monitor, all of the following are important in order to develop a quality control (QC) program,except:
A
Routine quality control tests by the QC technologist
B
Disassembly and cleaning of the internal monitor control devices by the QC technologist
C
Periodic review of the QC program by a qualified medical physicist
D
Annual and post-repair medical physics performance evaluations
BDisassembly and cleaning of the internal monitor control devices by the QC technologist
-
The image in Figure 6–6 was obtained while testing
A
rectifier operation in a single-phase x-ray machine.
B
rectifier operation in a three-phase x-ray machine.
C
timer accuracy in a single-phase x-ray machine.
D
timer accuracy in a three-phase x-ray machine.
 -
- D
-
- timer accuracy in a three-phase x-ray machine.
-
Which of the following may occur if the X-ray exposure field is not properly collimated, positioned, and sized?
A
Modulation transfer function failure
B
Moiré artifact
C
Exposure field recognition errors may occur
D
Ghost artifact
 -
- C
-
- Exposure field recognition errors may occu
-
Quality control testing of digital display monitors should be conducted
A
Daily
B
Weekly
C
Monthly
D
Annually
 -
- A
-
- Daily
-
Inadequate collimation in CR imaging can result in an image that is too
- light
- dark
- noisy
A
1 only
B
1 and 2 only
C
2 and 3 only
D
1, 2, and 3
 -
- D
-
- 1, 2, and 3
-
Which of the following is (are) tested as part of a QC program?
- Beam alignment
- Reproducibility
- Linearity
A
1 only
B
1 and 2 only
C
1 and 3 only
D
1, 2, and 3
 -
- D
-
- 1, 2, and 3
-
QA was being performed on a three-phase, full-wave-rectified x-ray unit. A synchronous spinning-top test was performed using 300 mA, 60 ms, and 70 kVp, and a 22-degree arc is observed on the test film. Which of the following statements regarding these results is most correct?
A
The timer is inaccurate.
B
The milliamperage station is inaccurate.
C
One rectifier is malfunctioning.
D
The test results are satisfactory.
 -
- D
-
- The test results are satisfactory.
-
Quality Control (QC) testing is used to evaluate digital display monitors in all of the following ways, except
A
Display luminance response
B
Display resolution
C
Display matrix size
D
Display noise
 -
- C
-
- Display matrix size
-
In order to avoid background radiation artifacts when using CR, it is important to:
A
Erase all image plates that have not been used for 48 hours
B
Erase all image plates if there is any question about how long it has been since the plate has been erased
C
Erase an image plate if there is any doubt as to when it was last erased, especially in the case of pediatric radiography
D
All of these are correct actions
 -
- D
-
- All of these are correct actions
-
Radiographs from a particular three-phase, full-wave-rectified x-ray unit, made using known correct exposures, were underexposed. A synchronous spinning top test was performed using 200 mA, 1/12 second, and 70 kVp, and a 20° arc is observed on the test film. Which of the following is mostlikely the problem?
A
The 1/12-second time station is inaccurate.
B
The 200-mA station is inaccurate.
C
A rectifier is not functioning.
D
The processor needs servicing.
 -
- A
-
- The 1/12-second time station is inaccurate.
|
|