Equipment Operation and Quality Control I

  1. To determine how quickly an x-ray tube will disperse its accumulated heat, the radiographer uses a(n)



    A
     
    technique chart. 


    B
     
    radiographic rating chart. 


    C
     
    anode cooling curve. 


    D
     
    spinning top test.
    •  
    • C
    •  
    • anode cooling curve.
  2. The ability of an x-ray unit to produce constant radiation output, at a given mA, using various combinations of mAs and time is called



    A
     
    linearity. 


    B
     
    reproducibility. 


    C
     
    densitometry. 


    D
     
    sensitometry.
    •  
    • A
    •  
    • linearity.
  3. In digital imaging, artifacts arise from a number of sources, including which of the following?



    A
     
    Imaging hardware 


    B
     
    Image processing software 


    C
     
    Operator error artifacts 


    D
     
    All of these may be sources of image artifacts
    •  
    • D
    •  
    • All of these may be sources of image artifacts
  4. All of the following are daily quality control (QC) steps to ensure optimal diagnostic images on a digital display monitor, except:



    A
     
    Turn on the monitor and allow it to warm up 


    B
     
    Evaluate luminance, reflection, noise and glare 


    C
     
    Make sure that the monitor is dust-free on the viewing surface 


    D
     
    Evaluate the QC patterns at the periphery and verify that all letters and numbers appear
    •  
    • D
    •  
    • Evaluate the QC patterns at the periphery and verify that all letters and numbers appear
  5. The x-ray beam and collimator light field must coincide to within



    A
     
    10% of the OID 


    B
     
    2% of the OID 


    C
     
    10% of the SID 


    D
     
    2% of the SID
    •  
    • D
    •  
    • 2% of the SID
  6. To compensate for variations in gain across a digital receptor, which of the following maintenance steps should be taken?



    A
     
    Conduct a calibration correction for image nonuniformity 


    B
     
    Increase or decrease the exposure factors to compensate 


    C
     
    Install a variable resistor to adjust the electrical supply to the unit 


    D
     
    Keep a log for at least 30 days to confirm consistent variations before making any adjustments
    •  
    • A
    •  
    • Conduct a calibration correction for image nonuniformity
  7. A device used to measure the luminance response and uniformity of monitors used in digital imaging is called a



    A
     
    Penetrometer 


    B
     
    Densitometer 


    C
     
    Sensitometer 


    D
     
    Photometer
    •  
    • D
    •  
    • Photometer
  8. Environmental background radiation can contribute to a



    A
     
    Ghost artifact 


    B
     
    Moiré artifact 


    C
     
    Static artifact 


    D
     
    Grid cutoff artifact
    •  
    • A
    •  
    • Ghost artifact
  9. Too much edge enhancement of the image in digital systems can cause an unwanted artifact called the:



    A
     
    Photoelectric effect artifact 


    B
     
    Scaling artifact 


    C
     
    Hounsfield artifact 


    D
     
    Halo effect artifact
    •  
    • D
    •  
    • Halo effect artifact
  10. All of the following are steps that should be used to accomplish quality control (QC) in digital radiography, except:



    A
     
    Acceptance testing 


    B
     
    Establishment of baseline performance 


    C
     
    Monitoring patient size to evaluate variations in equipment performance 


    D
     
    Diagnosis of changes in performance
    •  
    • C
    •  
    • Monitoring patient size to evaluate variations in equipment performance
  11. An exposed image plate will retain its original image quality for about



    A
     
    2 hours 


    B
     
    8 hours 


    C
     
    24 hours 


    D
     
    48 hours
    •  
    • B
    •  
    • 8 hours
  12. The device used to test the accuracy of the x-ray timer is the



    A
     
    densitometer 


    B
     
    sensitometer 


    C
     
    penetrometer 


    D
     
    spinning top
    •  
    • D
    •  
    • spinning top
  13. To ensure proper operation of the digital image display monitor, all of the following are important in order to develop a quality control (QC) program,except:



    A
     
    Routine quality control tests by the QC technologist 


    B
     
    Disassembly and cleaning of the internal monitor control devices by the QC technologist 


    C
     
    Periodic review of the QC program by a qualified medical physicist 


    D
     
    Annual and post-repair medical physics performance evaluations
    BDisassembly and cleaning of the internal monitor control devices by the QC technologist 


  14. The image in Figure 6–6 was obtained while testing







    A
     
    rectifier operation in a single-phase x-ray machine. 


    B
     
    rectifier operation in a three-phase x-ray machine. 


    C
     
    timer accuracy in a single-phase x-ray machine. 


    D
     
    timer accuracy in a three-phase x-ray machine.
    •  
    • D
    •  
    • timer accuracy in a three-phase x-ray machine.
  15. Which of the following may occur if the X-ray exposure field is not properly collimated, positioned, and sized?



    A
     
    Modulation transfer function failure 


    B
     
    Moiré artifact 


    C
     
    Exposure field recognition errors may occur 


    D
     
    Ghost artifact
    •  
    • C
    •  
    • Exposure field recognition errors may occu
  16. Quality control testing of digital display monitors should be conducted



    A
     
    Daily 


    B
     
    Weekly 


    C
     
    Monthly 


    D
     
    Annually
    •  
    • A
    •  
    • Daily
  17. Inadequate collimation in CR imaging can result in an image that is too

    1. light

    2. dark

    3. noisy





    A
     
    1 only 


    B
     
    1 and 2 only 


    C
     
    2 and 3 only 


    D
     
    1, 2, and 3
    •  
    • D
    •  
    • 1, 2, and 3
  18. Which of the following is (are) tested as part of a QC program?

    1. Beam alignment

    2. Reproducibility

    3. Linearity





    A
     
    1 only 


    B
     
    1 and 2 only 


    C
     
    1 and 3 only 


    D
     
    1, 2, and 3
    •  
    • D
    •  
    • 1, 2, and 3
  19. QA was being performed on a three-phase, full-wave-rectified x-ray unit. A synchronous spinning-top test was performed using 300 mA, 60 ms, and 70 kVp, and a 22-degree arc is observed on the test film. Which of the following statements regarding these results is most correct?



    A
     
    The timer is inaccurate. 


    B
     
    The milliamperage station is inaccurate. 


    C
     
    One rectifier is malfunctioning. 


    D
     
    The test results are satisfactory.
    •  
    • D
    •  
    • The test results are satisfactory.
  20. Quality Control (QC) testing is used to evaluate digital display monitors in all of the following ways, except



    A
     
    Display luminance response 


    B
     
    Display resolution 


    C
     
    Display matrix size 


    D
     
    Display noise
    •  
    • C
    •  
    • Display matrix size
  21. In order to avoid background radiation artifacts when using CR, it is important to:



    A
     
    Erase all image plates that have not been used for 48 hours 


    B
     
    Erase all image plates if there is any question about how long it has been since the plate has been erased 


    C
     
    Erase an image plate if there is any doubt as to when it was last erased, especially in the case of pediatric radiography 


    D
     
    All of these are correct actions
    •  
    • D
    •  
    • All of these are correct actions
  22. Radiographs from a particular three-phase, full-wave-rectified x-ray unit, made using known correct exposures, were underexposed. A synchronous spinning top test was performed using 200 mA, 1/12 second, and 70 kVp, and a 20° arc is observed on the test film. Which of the following is mostlikely the problem?



    A
     
    The 1/12-second time station is inaccurate. 


    B
     
    The 200-mA station is inaccurate. 


    C
     
    A rectifier is not functioning. 


    D
     
    The processor needs servicing.
    •  
    • A
    •  
    • The 1/12-second time station is inaccurate.