What is the difference between piston and pneumatic ventilators?
- Pneumatic requires a drive gas to drive bellows
- Piston system does not use a drive gas, so may be used on backup tank
-
At what pressures are positive and negative safety valves opened on piston ventilator?
- >75 +/-5cmH2O positive pressure relief valve opens
- < - 8cmH2O negative pressure relief valve opens
-
What does the negative pressure relief valve do on the piston ventilator?
- 1) prevents NEEP
- 2) allows room air to be drawn in
-
How is gas delivered in volume controlled mode?
- Driving gas flow stops at set tidal volume delivered to breathing circuit
- Or Pmax is reached
-
How is gas delivered in pressure controlled mode?
- Target pressure is set and maintained for certain time period
- The delivered tidal volume varies depending on compliance
-
What is the difference between ascending and descending bellows?
- Describes what the bellows do during EXPIRATION
- Descending have potential for NEEP
- Ascending always have some amount of PEEP due to weight of bellows
-
Why are ascending bellows safer than descending?
They do not fill if there is a disconnect
-
What is the function of the spill valve in the bellows?
- Opens if set pressure is exceeded to vent gas into WAG system
- It will open at any point if peak pressure is exceeded
-
At what pressure does spill valve open?
+2-3cmH2O
-
What is the most common course of action with ventilator alarms?
- Switch to manual ventilation
- Increase gas flow
- Check patient and system
-
Pressure below threshold for how long will signal apnea/disconnect alarm?
15-30 seconds
-
What can the reverse flow alarm indicate?
Incompetent expiratory valve
-
What is a good ventilator mode for difficulty ventilating?
PC-VG
-
What does PSV require?
Spontaneously breathing patient
-
What does PSVPro do in the event of apnea?
Switch to backup mode (SIMV on our machine)
-
In PSVPro backup mode, what happens when spontaneously breathing resumes?
Switches back to PSV mode
-
What are typical adult settings for VCV?
- TV 10ml/kg
- Rate 6-12
- PEEP 0 to start
-
What does the PIP alarm in VCV alert you to?
- Pt bucking
- Obstructed circuit
- Obstructed airway
-
What is the course of action for asynchronous breathing in VCV?
- 1) turn vent off
- 2) hand ventilate
- 3) remedy the situation
- Patients are almost always light
- If 1,2,3 doesn't work, disconnect circuit and ambu pt, call for help, switch to TIVA
- Still difficult, look at what is going on in the airway
-
What is the typical peak airway pressure alarm setting for relatively healthy adults in VCV?
40cmH2O
-
What types of patients is PCV indicated for?
LMAs, emphysema, neonates and infants
-
What types of patients have poor airway compliance and require higher volumes?
- Pregnancy
- Laparoscopic surgeries
- Morbid obesity
- ARDS
-
What are typical adult settings for PCV?
- 1) pressure limit 20cmH2O
- 2) rate 6-12
- 3) PEEP 0 to start
- This is also good to start for pediatrics
-
How is pressure limit affected by PEEP in PCV?
PC at 20, 5 of PEEP, inspiratory pressure is 25
-
What is less likely to occur in SIMV?
Asynchronous breathing
-
What is unique about SIMV?
- Detects spontaneous breaths (if any)
- Delivers controlled breaths in synch with the patient's inspiratory effort
- Helps maintain minute ventilation avoiding breath-stacking or bucking
-
What is the trigger window in SIMV?
Percentage of expiratory pause the vent monitors for patient effort
-
What is sensitivity in SIMV?
How much negative inspiratory force the pt must generate to trigger a breath (l/min)
-
How does PCV-VG work?
- Ventilator operates in PCV with a target tidal volume
- Dynamically adjusts inspiratory pressure until target tidal volume reached
- Does not exceed maxium set Pmax
-
What type of surgeries is PCV-VG good for?
Laparoscopic surgeries
-
What advantages does PCV-VG offer?
- Control of PIP - basic PCV mode
- Control of arterial oxygenation - guaranteed tidal volume
- Good default for most machines
-
What anesthesia considerations are there for ARPV?
Pt will remain on ICU vent and will do TIVA
-
What is the concept behind APRV?
- Vent cycles between 2 levels of CPAP
- High pressure and low pressure correlate to high volume and low volume
-
Where does gas exchange occur during APRV?
- When pressure is intermittently released to lower pressure
- 1) waste gases eliminated
- 2) oxygenation occurs
-
What is the normal I:E ratio?
1:2
-
How does I:E ratoi change in SIMV?
It changes as you change the rate
-
In what patients would an increased expiratory phase be beneficial?
- Obstructive lung disease
- 1) COPD
- 2) Emphysema
-
When is institution of PEEP recommended?
- 1) PaO2 cannot be kept >60mmHg
- 2) FiO2 at 50%
-
What are 3 hazards of PEEP?
- 1) decreased cardiac output
- 2) pulmonary barotrauma
- 3) increased extravascular lung water due to obstructed pulmonary lymph flow
-
Why is 4l/min FGF common?
From days when safety margin was needed for less accurate flowmeters and vaporizers
-
What FGF functions essentially the same as a nonrebreather mask?
- 5-8 lpm
- 1-1.5 times minute ventilation
-
When should high FGF (5-8lpm) be used?
- Preoxygenation
- Induction
- Emergence
-
When should low FGF (0.5-2lpm) be used?
During maintenance of anesthesia
-
What agent requires that you not use FGF <1 lpm for > 2 MAC hours?
Sevoflurane
-
What is a key disadvantage of low FGF?
- Accumulation of undesired gases in closed circuit
- Acetone, CO, methane, hydrogen, ethanol, anesthetic agent metabolites, argon, nitrogen
-
What is the most significant factor in promoting adequate denitrogenation?
Good mask seal
-
What is the most common site of disconnection?
Y-piece
-
What is the most common preventable equipment-related cause of mishaps?
Disconnection
-
What must be used to guard against disconnects?
Either ETCO2 or precordial stethascope
-
Can you silence apnea alarms?
No
-
What is the biggest problem with ventilators?
Failure to initiate ventilation or resume after pause
-
What is the most important monitor for disconnection?
Precoridal because it's alarms cannot be inactivated like ETCO2
|
|