It is a durable, reliable, easy-to-use portable instrument for detecting the presence of
Oxygen (O2),
Hydrogen Sulfide (H2S)
- Carbon Monoxide (CO)
- combustible gas, Lower Explosive Limits (LEL)
-
MSA Solaris Fire Multi-gas Detector – Description and Use
For detecting the presence of O2, H2S, CO and combustible gas
- Confined space entry
- CO alarm calls
- Natural gas leaks
- Gasoline spills
- Odor calls
- Overhaul
- Hazardous materials incidents
- Anytime a hazardous atmosphere is suspected
-
Additional Equipmen
Optional Sampling Pump and Tubing
3 AA Batteries
-
MSA Solaris Fire Multi-gas Detector
Specifications:
Rubberized armor unit is IP 65 rated for dust and liquid ingress.
Totally protects against dust.
Protects against water spray - not water proof!
- User-friendly
- Triple alarm system: audible, visual and vibrating.
- Auto-calibration
- Alkaline battery
- Weighs under 8oz
12 hours of continuous operation with 3 new AA batteries.
- 3 pushbuttons
- Intrinsically Safe
- UL913 certified for Class I, Div. 1, Groups A, B, C and D to ambient temperature of -20°F to +50°F
-
MSA Solaris Fire Multi-gas Detector
Alarm Levels:
O2 - 19.5% low and 23% high
H2S - 1 ppm
CO - 35 ppm
LEL - 10%
-
MSA Solaris Fire Multi-gas Detector
To turn on:
Press the "ON/OFF" button
Allow unit to perform self tests and self check (within 2 min)
-
MSA Solaris Fire Multi-gas Detector
Fresh Air Setup (FAS):
Once instrument self check is complete, ZERO? flashes for 10 seconds
To perform a Fresh Air Setup, push the ON/OFF button while ZERO? is flashing
-
MSA Solaris Fire Multi-gas Detector
To bypass the Fresh Air Setup:
-
Press RESET or do not push any buttons while ZERO? is flashing. After 10 seconds have expired the Fresh Air Setup will not be performed
-
MSA Solaris Fire Multi-gas Detector
Battery Warning:
Indicates a nominal 15 minute of operation remains.
- Battery life indicator flashes.
- Alarm sounds
- "BATT WRN" and lights flash every 15 seconds
- Leave atmosphere and replace batteries when a "Battery Warning" condition occurs
-
MSA Solaris Fire Multi-gas Detector
To turn off:
-
Press and hold "ON/OFF" button for approximately 3 seconds or until unit displays "OFF"
-
MSA Solaris Fire Multi-gas Detector
Sunday checks ensure the following:
Ensure battery life is greater than 50%
Ensure 3 additional AA batteries are available
- Ensure extension tube is present and functional
- Ensure pump is present and functional
-
MSA Solaris Fire Multi-gas Detector
Calibration:
Perform full calibration semi-annually.
DESIGNATED UNITS WILL BE NOTIFIED BY LOGISTICS.
-
MSA Solaris Fire Multi-gas Detector
Send unit to Logistics to recheck calibration sooner if any of the following occurs:
High impact physical shock
Exposure to high levels of contaminants
Exposure to lead-containing compounds
Exposure to Hydrogen Sulfide (H2S) 50 ppm for more than one minute or any exposure over 200 ppm
|
|