-
Galvanic cell
- •
- Measures the current produced when O2 diffuses across a membrane &
- it's reduced to molecular oxygen at the anode of the electrical circuit
- •
- The current is proportional to the ppO2 in the fuel cell
- •
- These analyzers require regular replacement of the sensory capsule
- The elec. potential for reduction of O2 results from a chemical reaction and over
- time, the reactants require replacement
-
Paramagnetic analysis
- ·
- : Oxygen
- is a nonpolar gas, but it is paramagnetic and when placed in a magnetic field,
- the gas will expand, contracting when the magnet is turned off. By switching
- the field on and off and comparing the resulting change in volume (or pressure
- or flow) to a known standard, the amount of oxygen can be measured.
-
Polargraphic electrode:
· has a gold (or platinum) cathode and a silver anode, both based in an electrolyte, separated from the gas to be measured by a semipermeable membrane. Unlike the galvanic cell, a polarographic electrode works only if a small voltage is applied to two electrodes. The amount of current that flows is proportional to the amount of oxygen present."
-
Laburt beer law:
oximetry).
- ·
- Oximetry depends on the observation that
- oxygenated and reduced hemoglobin differ in their absorption of red and
- infrared light (Lambert–Beer law).
- Specifically, oxyhemoglobin (HbO2) absorbs more infrared light (960 nm),
- whereas deoxyhemoglobin absorbs more red light (660 nm) and thus appears blue,
- or cyanotic, to the naked eye. The change in light absorption during arterial
- pulsations is the basis of oximetric determinations (Figure 6–22). The ratio of
- the absorptions at the red and infrared wavelengths is analyzed by a
- microprocessor to provide the oxygen saturation (SpO2) of arterial blood.
- Arterial pulsations are identified by plethysmography, allowing corrections for
- light absorption by nonpulsating venous blood and tissue.
-
Limiting factors of pulse ox
- ·
- Because carboxyhemoglobin (COHb) and HbO2 absorb
- light at 660 nm identically, pulse oximeters that compare only two wavelengths
- of light will register a falsely high reading in patients with carbon monoxide
- poisoning.
- ·
- Methemoglobinemia causes a falsely low
- saturation reading when SaO2 is actually greater than 85% and a falsely high
- reading if SaO2 is actually less than 85%.
- ·
- Bronchial intubation will usually go undetected
- by pulse oximetry in the absence of lung disease or low fraction of inspired
- oxygen concentrations (FIO2).
- ·
- Other causes of pulse oximetry artifact include
- excessive ambient light, motion, methylene blue dye, venous pulsations in a
- dependent limb, low perfusion (eg, low cardiac output, profound anemia,
- hypothermia, increased systemic vascular resistance), malpositioned sensor, and
- leakage of light from the light-emitting diode to the photodiode, bypassing the
- arterial bed (optical shunting).
-
What can cause increases in exhaled concentrations of CO2?
- ·
- Increases in Exhaled CO2 can occur from:
- o
- Increased CO2 production
- o
- CO2 rebreathing: Exhausted absorber, faulty
- unidirectional valves
- ·
- Decreases in Exhaled CO2 occur from:
- o
- Pulmonary hypoperfusion: Decreased cardiac output,
- PE, congenital right to left shunt
- o
- Leaking shallow breathing
- o
- Rapid shallow breathing
- ·
- End-expired CO2 absent:
- o
- ETT misplacement: extubated, esophageal
- o
- Complete tube obstruction
-
What is Aspiration Capnography?
- ·
- Diverting (Aspiration) Capnograph: continuously
- suctions gas from the breathing circuit into a sample cell within the monitor
- o
- Prone to water precipitation into ETT
- ·
- VERSUS: Nondiverting (Flowthrough): measures CO2
- passing through an adaptor placed in the breathing circuit
-
Circumference of BP cuff, what happens if to big or small
- ·
- Cuff/arm relationship: the cuff’s bladder should
- extend at least halfway around the extremity, and the width of the cuff should
- be 20% to 50% greater than the diameter of the extremity
- o
- Influences the pressures measured
- o
- Based on thickness of extremity, not length
- o
- More peripheral= SBP increases and DBP decreases
- o
- 10cm above/below heart= 7.5 mm Hg added to or
- subtracted
- ·
- FALSE HIGH = cuff too small
- ·
- FALSE LOW = cuff too big
-
Complications of arterial line
- ·
- Hematoma/bleeding-catheter separation
- ·
- Embolization of air bubbles/thrombi
- ·
- Necrosis of skin overlying catheter
- ·
- Unintentional intra-arterial drug injection
-
Who
benefits from a PA-line
- CAD with LV dysfunction,
- recent infarct – valvular heart disease – CHF
- Severe COP – acute
- respiratory failure
- Shock – acute renal failure –
- burns – pancreatitis
- Pericardiectomy –
- aortic-cross clamping – sitting craniotomy – portal systemic shunt – liver
- transplant
- Severe toxemia – placental
- absorption
-
Causes
of hypothermia
- Ambient temperature below 21
- degree Celsius
- Anesthetic-induced
- interference with the hypothalamic thermostat
Vasodilation
- Decreases in basal metabolic
- rate
-
Delta rhythm on EEG
- (0-3
- Hz):
- deep sleep, deep anesthesia, or pathologic states (brain tumors, hypoxia,
- metabolic encephalopathy) –
- very low frequency due to ischemia
-
Theta
- (4-7 Hz): sleep and anesthesia in adults,
- hyperventilation in awake children and young adults – cortical depression
-
Alpha EEG rhythm
- (8-13 Hz): most common normal rhythm –
- steady wavelength seen in resting,
- awake adult with eyes closed
-
Beta EEG rhythm
- (>13 Hz): when alpha rhythm is disrupted by
- excitation, a less synchronous higher frequency beta rhythm is
- seen – mental activity,
- light anesthesia
-
. BIS monitor numbers & range
- To perform a BIS analysis, data measured from the EEG is
- calculated to a single number that correlates with the depth of
- anesthesia.
- • This
- value from 65-85 has been advocated as a measure of sedation,
- • Values
- of 40-65 have been recommended for GA.
-
what the BIS monitor
IS, and also what it is NOT.
- Optimal use requires
- knowing what the BIS monitor IS, and also what it is NOT.
- • Particularly, the BIS monitor is NOT a
- MAC-meter (It cannot tell you if the patient is at MAC or not).
- • It
- does NOT predict the likelihood of movement in response to an incision (or any
- noxious stimuli for that matter).
- • The
- patient could be 40 on the BIS monitor, but if you have not given enough
- narcotic, the patient can still move.
- • Cerebral
- ischemia can produce a loss of consciousness, and the BIS monitor may
- consequently decrease
- • But,
- significant cerebral ischemia can occur without noticeable difference in our
- BIS monitor
- • BIS
- is a form of EEG, but we would not use it instead of a real EEG to do a carotid
- endarterectomy
- • Can
- be used with a propofol drip, for example, to make sure the patient is asleep
-
Evoked Potentials
- ·
- Diagnostic tool to evaluate certain neurologic
- disorders
- ·
- Monitor functional integrity of sensory and motor pathways during
- surgical procedures
- o
- Spinal fusion with instrumentation
- o
- Spine and spinal cord tumor resection
- o
- Thoracoabdominal aortic aneurysm repair
- o
- Cerebral tumor resection
- ·
- Extremely small amplitude (microvolts) electrical
- potentials generated by nervous tissue in response to stimulation
-
·
Brainstem
auditory evoked responses (BAERs)
- o
- Monitor by stimulating the cochlea
- o
- Useful in assessing brainstem function in
- comatose patients
- o
- Surgical procedures of the cerebellopontine
- angle, floor of the fourth ventricle, or fifth, seventh, or eighth cranial
- nerve
- o
- Resistant to effects of anesthesia
-
·
Visual
evoked potentials (VEPs)
- o
- Produced by flashing light to stimulate the
- retina
- o
- Records EPs over the occipital cortex
- o
- Assess the integrity of the visual pathway
- o
- Used in resection of pituitary tumors,
- craniopharyngiomas, or near optic tracts
-
·
Somatosensory
evoked potentials (SSEPs)
- o
- Monitors transmission of EPs through the sensory
- pathway
- o
- Nerves are median, ulnar, peroneal, or posterior
- tibial
- o
- Monitor cerebral function or ischemia
- o
- Evaluate spinal cord function
-
·
Motor
evoked potentials (MEPs)
- o
- Assesses descending motor pathways during
- neurosurgical, orthopedic, or vascular procedures
- o
- Obtained by transcranial electrical or magnetic
- stimulation or direct spinal cord stimulation
-
Equation for air and oxygen and how much is being
delivered:
- If you are delivering 1 L of O2 and 1 L of air, what % O2 is
- being delivered?
- So there is 1 L of 100% O2 and 1 L of 21% O2.
- 100 + 21= 121
- 121 divided by 2= 60% O2
- You should also know how to do this with N2O
- (remember, there is no O2 in N2O).
-
What are the characteristics of a laser
- o
- Monochromatic- possesses one wavelength
- o
- Coherent- it oscillates in the same phase
- o
- Collimated- exists as a narrow, parallel beam
-
o
Uses & side effects of lasers:
- ·
- Depends on wavelength, which is determined by
- the medium in which the laser beam is generated.
- ·
- A laser medium is a substance that can be
- stimulated to emit a stable state when pumped with an external energy
- source.
- ·
- Can be solid, gas, liquid, or
- semiconductor. An example of this is
- that a medium of CO2 gas laser produces a long wavelength laser while a Yag laser (which is
- solid-state laser) results in a shorter wavelength.
- ·
- The longer the wavelength, the greater the
- absorption of water, and the less tissue is penetrated. Therefore, the CO2 laser’s effects are much
- more localized & superficial than the Yag laser. However, the solid-state lasers such as the
- Yag laser are more powerful than gas lasers.
o
-
What is the fire triangle? What are its
components?
- The fire triangle is depicted below. It
- demonstrates that no one person alone is responsible for fire prevention. Fire prevention requires intraoperative
- collaboration with at least one of the other two professions.
- ·
- The surgeon
- yields the ignition source.
- ·
- Anesthesia
- controls the oxidizers (oxygen and
- nitrous).
- ·
- RN
- personnel are frequently in control of the safe use of potential fuels (alcohol prepping solutions and
- sponges).
-
What is an oxidizer?
- ·
- An oxidizer is a substance that will support the combustion of fuel. Most
- fuels burn only in the gaseous state and ignite only when sufficient vapors mix
- with oxygen. Heat produces these
- vapors by evaporating liquids or vaporizing solids. Although oxygen from the
- air combines with the fuels during surgery during a fire, the OR has other
- sources of oxygen.
- ·
- Anesthesia requires the delivery of oxygen concentrations
- above 21%. Whenever and wherever the O2 concentration is above 21% = O2
- enriched environment. With increased O2 a fire is easy to ignite, it will burn
- faster and hotter, and it will be difficult to extinguish.
- ·
- Oxygen is supplied from the anesthesia machine,
- which is a ventilator, wall outlet, or the gas cylinder.
- ·
- O2 is heavier than air, it will collect in
- low-lying cavities/areas: open chest cavity or under the drapes.
- ·
- Drape fabrics will absorb oxygen and retain it
- for some time.
- ·
- Oxygen-enriched environment will also allow some
- materials (ie. plastic) to burn that would otherwise not normally burn at RA.
- ·
- Oxygen for a fire can also be supplied from
- thermal decomposition of N2O. Heat
- from sources found in OR or fire will liberate oxygen from nitrous, allowing it
- to support combustion.
- ·
- Within the context of surgical fires, any
- mixture of oxygen and nitrous is considered to be an oxygen-enriched
- environment.
- ·
- In the event of a fire, the supply of nitrous as
- well as oxygen must be shut off quickly.
- ·
- OR supplies that support combustion: ETT, O2
- catheters, surgical drapes, benzoin, alcohol cleansing solution,
- petroleum-based ointments (lacrilube)
-
What is a fuel?
- ·
- Fuel is
- anything that can burn – Including almost anything that comes in contact
- with patients as well as the patients themselves.
- ·
- Some more commonly found fuels in surgery
- include hair, GI tract gases (methane, hydrogen, hydrogen sulfide), drapes,
- gowns, sponges, breathing circuits, etc.
- ·
- CDC recommends alcohol-based gel solutions for
- hand hygiene and alcohol-based chlorhexidine solutions for skin preps. Now they
- say the patient cannot be draped for 3 minutes – RNs should also ensure
- that there is no pooling of solution.
- ·
- Potential for fire is augmented when the
- alcohol-based antiseptic is applied in ways that allows pooling (may not even
- see it because it under the patient). Sponges themselves will help to prevent
- that.
- ·
- Providone-iodine solutions are also flammable.
- Similar cautions must be made during prepping.
- ·
- Under the right conditions, some surgical
- ointments can burn. Some petroleum-based ointments used in oxygen-enriched
- environments will ignite when enough heat is present to cause vaporization.
- These materials must vaporize and mix with oxygen to allow ignition. Globs will
- not easily ignite because its mass absorbs considerable heat before
- vaporization occurs. However, thin layers have low mass per area and need less
- heat to cause vaporization, and are easier to ignite.
- ·
- Water-based ointments (surgilube) will not burn
- easily. Head and neck surgery – may put surgilube in the patient’s hair.
-
What is an ignition source?
- ·
- Heat
- input from a variety of sources increases the oxidation rate of fuel, oxygen
- mixture until combustion will occur.
- ·
- In addition to the overhead surgical lights,
- some of the other heat sources found in the OR include defibrillators, cautery
- units, heated probes, drills & burrs, fiberoptic light sources & the
- cables, and lasers. These sources produce temperatures from several hundred to
- a few thousand degrees Fahrenheit. (Enough to ignite most fuels.)
- ·
- Incandescent sparks produced by electric cautery
- or high-speed drills and burrs can cause fires.
- ·
- Lasers can case sparks when energy hits an
- instrument or the laser fiber itself may be damaged.
- ·
- Even glowing embers of charred tissue can
- provide enough initial heat to ignite some fuels – especially when you are
- using oxygen around it.
- ·
- A few seconds after deactivation, a heated
- cautery probe tip, a fiber optic cable tip, or a laser contact tip can retain
- enough heat to melt plastics or ignite fuels.
- ·
- While these devices must be in contact with the
- material to heat it, a laser can heat from a few centimeters to several meters
- away. A fiber optic light source may take a minute or so to heat a drape to the
- point of combustion, while a laser can cause this almost instantaneously.
- ·
- By ensuring that these heat sources are not
- directed towards or allowed to come in contact with the fuel, the OR staff can
- prevent these fires.
- ·
- Electrical equipment: electrosurgical unit
- (cautery near distended bowel), lasers (used near ETT)
- 54. safety devices on
- anesthesia machine that prevent anesthetic agent overdoses
Capnography and anesthetic gas measurement
-
Control various heat sources and prevent them from
contacting fuels
n Vigilance
- n Allow
- surgical preps to dry
- n Minimize
- oxygen concentration
-
Requisites for a fire or explosion
1. Flammable agent (fuel)
- b. bowel gas:
- methane, hydrogen, hydrogen sulfide
2. A gas that supports combustion
- a.
- operating room supplies: o2 catheters, surgical drapes, petroleum based
- ointments, petroleum based lacrilube
- b. oxygen
- and nitrous oxide
3. Source of ignition
- a. static
- electricity: decrease by maintaining relative humidity
- b. electrical equipment:electrical surgical
- units, cautery, laser, distended bowel, laser near ETT
-
. protecting the patients eyes during a laser surgery
Patients eyes should be taped and operating room personel should wear protective eyeglasses
-
Difference between CO2 laser and Yag laser? Wavelengths? What they do?
- LASER =
- Light amplification by stimulated emission of radiation
- ·
- Lasers different from normal light in 3 ways:
- o
- Monochromatic- possesses one wave-length
- o
- Coherent- it oscillates in the same phase
- o
- Collimated- exists as a narrow, parallel beam
- ·
- LASER
- = Light amplification by stimulated emission of radiation
- o
- Use and s/e of a laser depends on wave length,
- which is determined by the medium in which the laser beam is generated
- o
- Laser medium is substance that can be stimulated
- to a meta stable state when pumped with external energy source- either a solid,
- gas, liquid or semiconductor
- §
- Medium CO2 gas laser produces a long wavelength laser while a YAG
- laser, which is a solid-state laser, results in a shorter wavelength.
- §
- The longer the wavelength the greater the absorption of
- water and less tissue is penetrated. Therefore, the CO2 laser effects are much
- more localized and superficial than YAG.
- Shorter wavelength lasers including the YAG are more powerful but will
- hit a larger area.
-
· Common laser hazards:
- o
- Thermal
- trauma- misdirected beams potential risk for patient and OR staff- why we wear
- goggles too, skin trauma dependent on energy and wavelength of laser, varies
- from reddening to blistering and charring
- o
- Eye
- injury- see question 19 for eye safety, eye is especially at risk because
- lasers radiation is coherent and all energy can be focused on very small
- portion of cornea or retina: CO2 injures cornea YAG injures retina, NO
- petroleum based ointments during laser procedures because they can cause burns
- o
- Perforation
- of organs or vessels- usually d/t misdirected laser beams, can result in
- bleeding or edema that may not appear for hours or days after surgery
- o
- Gas
- embolism- rare but potentially fatal
- Pollution- lasers create smoke with toxic gases
- and vapors, CO2 laser causes most smoke, high concentrations of smoke cause
- ocular and upper respiratory irritation, some lasers may be used with portable
- smoke evacuators, special filtration masks may be required for staff
- o
- during
- certain procedures to filter out airborne particles and airborne particles
-
Safety precautions when using a laser in the airway
- Endotracheal
- tubes support combustion-some are more resistant that others but there isn’t
- one that prevents it all together
- Polyvinyl
- is what is typically used
- Also
- a metallic tape that can be used- would need to prep ETT in AM/before surgery
-
Laser Safety Precautions:
· Inspired O2 kept low
- o
- Need
- to change O2 analyzer to allow you to administer <30%
- o
- Make
- sure your anesthesia machines has air hooked up to it
- · Nitrous Oxide NOT used- nitrous
- supports combustion
· ETT cuff filled with fluid
- o
- Fill
- cuff with saline dyed with methylene blue so can recognize if cuff has been
- pierced
- · Laser intensity and duration
- limited
· Saline-soaked pledgets in airway
-
What actions to take if a fire occurs in the airway?
- Don’t remove lift drapes into air
- re can fuel fire-pull off horizontally! TAKE OUT ETT AND REINTUBATE!
-
What do we do for patients with a pace maker to prevent burns and
so the pacemaker doesn’t go off with use of cautery?
- · Micro-shock-results from current
- being supplied to externalized conductor such as saline filled catheter or
- packing wire BUT only effects patients that are electrically susceptible ie pt
- with CVL or pacemaker, results in internal injury
- o
- Very
- small amount of current can cause damage 100 microamp = vfib (unlike with macro need 100-300 milliamp)
- o
- Total
- leakage of OR equipment must be under 10 microamps to keep patient from being
- affected
- · Cautery- very frequently used by
- surgeons for cutting and tissue coagulation
- o
- Generate
- ultra high frequency current that passes from cautery tip through patient and
- exits via grounding pad
- o
- May
- cause shock, burns, explosions, arrhythmias, and disturbance in pacemaker
- function
- § Smoke inhalation- unknown if its
- harmful but recommended that it be drawn out with suction
- § Extensive tissue burns- patients
- are wet with blood and/or fluid and make electrical contacts with electrically
- conductive materials including OR table, monitoring electrodes, surgical
- retractors, stirrups = DANGEROUS PATHWAY FOR CURRENT won’t be shocked re
- transformer but can be burned if touching one of these electrically conductive
- materials/metal
- o
- GROUNDING
- PAD (Bovie)- to return high current from unit to low current and return it back
- to cautery unit,
- Correct
- placement = current dispersed over large area
- If grounding pad buckles energy
- will go to small area and cause burn
- ·
- 2 types of cautery: unipolar and bipolar
- §
- Most commonly used- to cauterize vessels
- §
- Grounding pad placed on patient- if have to use
- with pacemaker make sure pad is not near pacemaker, place it as far away as
- possible, same idea if patient has any metal in body
- § Used
- for less vascular tissue,
- §
- Used especially for patients with AICD (automatic implantable
- cardioverter defibrillator) or pacemaker re current will stay LOCALIZED, will
- only go from cautery pen and back to cautery unit
Principles of Electricity The Ohm’s law we know
Ohm’s Law: E=I x R BP = CO x SVR
- E= electromotive
- forces (volts) BP = volts
I= current (amperes) CO = current
- R= resistance (ohms) SVR
- = resistance to the forces opposing the flow of electrons
- Electrical devices in surgery have the potential to
- cause electrocution and burns in patients and OR staff. It is our
- responsibility to have some basic knowledge.
- Ohm’s law correlates the flow of electricity, the
- applied electrical pressure and the resistance to this flow.
- In order for electricity to occur electrons must move
- from an area of high concentrations to low area of concentration. A potential difference
- needs to exist between these two points. (Expressed in volts)
-
Micro-shock
- · Micro-shock-results from current
- being supplied to externalized conductor such as saline filled catheter or
- packing wire BUT only effects patients that are electrically susceptible ie pt
- with CVL or pacemaker, results in internal injury
- o
- Very
- small amount of current can cause damage 100 microamp = vfib (unlike with macro need 100-300 milliamp)
- o
- Total
- leakage of OR equipment must be under 10 microamps to keep patient from being
- affected
-
Cautery
- · - very frequently used by
- surgeons for cutting and tissue coagulation
- o
- Generate
- ultra high frequency current that passes from cautery tip through patient and
- exits via grounding pad
- o
- May
- cause shock, burns, explosions, arrhythmias, and disturbance in pacemaker
- function
- § Smoke inhalation- unknown if its
- harmful but recommended that it be drawn out with suction
- § Extensive tissue burns- patients
- are wet with blood and/or fluid and make electrical contacts with electrically
- conductive materials including OR table, monitoring electrodes, surgical
- retractors, stirrups = DANGEROUS PATHWAY FOR CURRENT won’t be shocked re
- transformer but can be burned if touching one of these electrically conductive
- materials/metal
-
GROUNDING PAD
- o
- GROUNDING
- PAD (Bovie)- to return high current from unit to low current and return it back
- to cautery unit,
- Correct
- placement = current dispersed over large area
-
·
2 types of cautery: unipolar and bipolar
Unipolar
- §
- Most commonly used- to cauterize vessels
- §
- Grounding pad placed on patient- if have to use
- with pacemaker make sure pad is not near pacemaker, place it as far away as
- possible, same idea if patient has any metal in body
- § Used
- for less vascular tissue,
- §
- Used especially for patients with AICD (automatic implantable
- cardioverter defibrillator) or pacemaker re current will stay LOCALIZED, will
- only go from cautery pen and back to cautery unit
-
Principles of Electricity
Principles of Electricity The Ohm’s law we know
Ohm’s Law: E=I x R BP = CO x SVR
- E= electromotive
- forces (volts) BP = volts
I= current (amperes) CO = current
- R= resistance (ohms) SVR
- = resistance to the forces opposing the flow of electrons
- Electrical devices in surgery have the potential to
- cause electrocution and burns in patients and OR staff. It is our
- responsibility to have some basic knowledge.
- Ohm’s law correlates the flow of electricity, the
- applied electrical pressure and the resistance to this flow.
- In order for electricity to occur electrons must move
- from an area of high concentrations to low area of concentration. A potential difference
- needs to exist between these two points. (Expressed in volts)
-
Know difference
between direct current and alternating current
DIRECT CURRENT (DC):
- ·
- Direct flow of electrons in the same direction.
- (ex: Flashlight battery, laryngoscopes, twitch monitors)
ALTERNATING CURRENT (AC):
- ·
- Alternating flow of electricrons occurs in 1
- direction then reverses itself at regular intervals. Electrical Power Company. In OR, most
- current is AC.
- ·
- In the US, utility companies supply electrical energy
- in the form of ACs of 120volts at a frequency of 60Hz.
- *Either
- of these types of current can be pulsed or continuous in nature.
- Ohm’s law is accurate when applied to DC circuits. But with
- AC circuits, the situation is more complex because the flow of the current is
- opposed by more complicated form of resistance called impedance. Impedance is
- the sum of the forces that oppose electron movement in the AC circuit. It
- consists of resistance (Ohm’s) but also takes capacitance and inductance into
- account.
-
whats a hertz
- Hertz = frequency in cycles per second at which the AC
- current reverses direction ( # of times the AC reverses itself in 1 second). 60
- hertz in USA. As frequency increases,
- impedance to flow decreases and more current is allowed to pass.
-
VOLT:
- The SI unit of electromotive force, the difference of
- potential that would carry one ampere of current against one ohm resistance
-
AMPERE
flow of electrons/sec past a given point. When this happens, heat & light are produced
-
RESISTANCE:
- Resistance includes forces that oppose current flow or is an impendence to flow. Higher resistance, less amount of current will flow
- If short circuit, there will be 0 impedance.
-
. what is the
electrical power in the OR
- ·
- Whereas electrical
- power is grounded in the home, it is usually UNGROUNDED in the OR.
- ·
- In the home, electrical
- equipment may be grounded or ungrounded but it should ALWAYS BE GROUNDED in
- the OR.
- ·
- In a grounded power system, it is possible to
- have either grounded or ungrounded equipment, depending on when the wiring was
- installed and whether the electrical device is equipped with a three prong plus
- containing a ground wire.
- ·
- Electrical
- contacts w/ground can cause injury when they complete the circuit ( that
- permits the pt completes it) but now we isolate the electrical power instead of
- being grounded, it is UNGROUNDED POWER
- ·
- In the OR, numerous electrical devices, together
- w/power cords and puddles of saline solutions on the floor, make an
- electrically hazardous environment for both patients and personnel.
- ·
- To
- provide an extra measure of safety from macroshock, the power supplied to most
- ORs is ungrounded!!
- ·
- The 120 volt potential difference exists only
- between the two wires of the isolated power system (IPS) but no circuit exists
- between the ground and either of the isolated power lines.
-
what is your line isolation
monitor and what does it protect against, micro/macro shock.
- · The isolation transformer allows the
- electrical power to be ungrounded in the OR. The electrical power in the OR
- comes from a hospital source, the usually originates from a connection to an AC
- from the local power company. After arrival to the OR, electrical power is
- modulated, isolated, and dispensed to electrical outlet in the room by
- secondary coils. Here, you will either have one or more large isolation
- transformers. Each isolation transformer is required to have a line isolation
- monitor, which is a simple electrical current meter that demonstrates the
- isolation of the transformer’s out power from the ground. If there is a short
- circuit, you will see the monitor fluctuate, and it will shut off the power.
- The line isolation monitor verifies that the power lines from the transformer
- are indeed isolated from the ground. It is connected to both sides of the
- isolated power outlet (the primary and secondary coils) and is set to alarm and
- shut off the power when either side has an impedance to the ground of less than
- 25,000 ohms, or when the max current reaches 2 milliamps from a short circuit.
- So the line isolation monitor is insensitive to currents below 2 milli amps.
- Therefor it will provide not protection against Micro Shock, it only protects
- against macro shock.
-
what are common causes of short
circuits in OR.
- · Dripping saline or water onto the
- extension cords or outlets,
-
Know volts of micro and macro
shocks
- · Micro shock- Result when the electrical
- current is accidently applied to an external conductor such as a saline filled
- catheter, or an electrical pacing wire. This only concerns patients that are
- electrically vulnerable: i.e. patients with a pacemaker, central lines.
- Measured in Micro ampules. Very small amounts of current can cause damage. So in
- order to prevent this, the total leakage of OR equipment can only be 10 micro
- amps.
- o
- 100
- micro amps – Causes V Fib
- · Macro shock- describes the effects of
- current applied to the body through intact skin. The flow of current takes the
- path of least resistance. Which would be the great vessels, nerves, and
- muscles. The units of measure associated with macro shock are milli ampules.
- Occurs from currents that rise from equipment failure, unsafe design, or
- misuse.
- o
- 1
- milli amp (mA) – Perception
- o
- 5 mA
- – Max harmless current
- o
- 10-20
- mA – “let go” current
- o
- 50 mA
- – loss of consciousness
- o
- 100-300
- milli amps – V Fib, respiratory center
- remains intact
- o
- 6000
- mA – complete physiologic damage
-
What should be the
total leakage of current in the OR, electrical current from equipment?
- Kathy said to remember this number: 100-300 milliamps (macroshock) will cause V
- Fib, respiratory center intact versus 100 microamps
- to cause V Fib
- In order to prevent microshock, the total leakage of OR
- equipment must be 10 microamps, so if
- we have a little current going on, as long as it’s under 10 microamps the patient will not be
- affected
-
Causes of fires in
the OR like instruments
- 74% of all cases of surgical fires have been in an oxygen
- enriched environment, a common contributor during head and neck surgery is the
- delivery of large concentrations of oxygen via a facemask
- Drapes covering the patient result in accumulation of
- concentrated oxygen under them, referred to as tenting; therefore, in an O2
- rich environment, application of an ignition source (such as the electrocautery unit, defibrillator, a hot fiber optic light
- source, or surgical laser) can regularly ignite this
Surgeon wields ignition source
- Static electricity in the past hospitals would try to
- decrease static electricity by maintaining humidity at 50%, gets to grade point
- where instruments aren’t sterile anymore
- Electrical equipment that is found in the OR- includes
- electrosurgical units (cautery), lasers
-
EPA
- nvolved in disinfection and sterilization of devices and
- procedures
-
FDA
- : regulates chemical germicides used on medical devices,
- also requires a manufacturer of reusable product to provide adequate
- instructions on cleaning and disinfecting
-
OSHA:
- regulates occupational exposure to chemical disinfectants
- and sterilizers
-
CDC
- recommends broad strategies to prevent transmission of an
- infection in the health care environment
-
what are the most resistant types of germs? Blood borne pathogens?
- Most resistant type
- of germs: bacterial spores (some of which are resistant to chemical and
- physical stresses) Hep B and HIV are the least resistant to chemicals
-
Antiseptic
- a substance that may be applied to living tissue, it has
- antimicrobial activity
-
Bacteriostat
- agent that prevents bacteria growth but it will not kill the
- bacteria
-
Decontamination-
- process by which contaminated items are rendered safe for
- personnel who are not wearing protective attire
- A.
- Reasonably free of (probably) transmitting infection
- B.
- Reduction of microbial contamination to an acceptable level
- C.
- Any process that Eliminates harmful
- substances
-
Disinfectant
- chemical germicide that is formulated to be used solely on
- inanimate objects (NOT HUMANS!)
3 Types of Disinfectants
- 1) High level- kills all organisms
- including bacterial spores and certain viruses; most of these can produce
- sterilization with sufficient contact time
- 2) Intermediate Level- kills
- bacteria including TB, some fungi, most viruses, but NOT bacterial spores
- 3) Low Level- kill most bacteria,
- NOT TB, some fungi, viruses, and spores
-
Sterilization
destruction of all viable forms of microorganisms
- 6 methods of sterilization “that
- you should be familiar with”
- 1) Pasteurization- equipment is
- immersed in water at an elevated temp for a given period of time. It is a
- disinfecting process that cannot be depended on for sterilization ex.
- ventilator bellows, laryngoscope blades (first step in sterilization process,
- this isn’t the only way blades are cleaned)
- Advantage to this method: lower
- temperatures: there’s less damage to the equipment and also no toxic fumes or
- residues
- 2) Steam Sterilization-aka:
- autoclaving: uses saturated steam under pressure, it will kill all bacteria,
- spores, and viruses; this is the method that is used in OR to clean ALL
- surgical equipment
- 3) Chemical Disinfection and
- Sterilization- utilizes liquid chemical agents and is often performed by
- soaking an item in the solution; it is useful for heat sensitive equipment
- Disadvantage: chemicals can be
- absorbed into the items and cause harm to the patients, cannot be used for all
- types of equipment, sterility is not guaranteed, some solutions aren’t safe for
- tissue, will have an unpleasant odor and you will need to avoid prolonged skin
- contact or inhalation of these vapors
- 4) Gas Sterilization- kills
- bacteria, spores, fungi, and viruses, it is flammable, and is a more complex
- and extensive process, it is restricted to objects that might be damaged by
- heat or excessive moisture, other complications: due to residual left on
- sterilized items that may cause skin reactions or laryngeal/tracheal
- inflammation- ethylene oxide
- 5) Radiation Sterilization- gamma
- rays; used for sterilizing disposable products from manufacturers, products are
- pre-packaged before treatment and will remain sterile indefinitely until
- package is opened (ex. ETT)
- 6) Gas Plasma Sterilization- uses
- gaseous chemical germicide and gaseous plasma, it is used on packaging
- materials, plastic, and stainless steel instruments and the one you’re going to
- hear is called the Sterrad system
-
Not a single chemical germicide that is adequate for all
purposes, some of the more common ones: alcohol, iodophores, or cydex
- 1.
- Alcohol- intermediate level germicide, it will
- kill most bacteria, but not spores, the effect of this is limited due to the
- rapid vaporization and it’s also flammable; ex chlorohexidine
- 2.
- Iodophores- combination of iodine and
- solubilizing agent; principally used at antiseptic
- 3.
- Cydex- high level disinfectant that is used in a
- 1% concentration and it MUST be rinsed thoroughly because it’s a physical
- irritant and will cause patient issues
-
difference between
uni polar and bipolar cautery and why we use one or another
unipolar
unipolar
-most commonly used in the OR
- current flows thru pt, to a grounding pad, then back to
- cautery unit itself
-used to cauterize vessels
- -if using this type with a pt that has a pacemaker, place
- bovie pad away from pacemaker (on thigh for example)
- -can’t use the grounding pad if the patient has metal in
- their body
bipolar
-used for less vascular tissues
-no grounding pad used with this type
-recommended for a pt with an AICD or PM
-current is localized and does not go thru pt’s body
-current goes from cautery pen back to the unit
-
what are the monitors
we check during pre anesthesia morning checklist
capnometer
-O2 analyzer
-pressure monitor with high and low airway pressure alarms
-pulse ox
-respiratory volume monitor (spirometer)
-
what is on the
anesthesia checklist
- emergency ventilation equipment (ambu bag and mask,
- intubating stylets)
- -high pressure system (check O2 cylinder supply- should be
- at least 1,000 psi, check central pipeline pressures- should be 50-55 psi)
- -low pressure system (check vaporizer levels and fill if
- needed, test flowmeters- attempt to create a hypoxic mixture with O2 and N20)
- -check scavenging system (check connections and adjust waste
- gas vacuum- bobbin should be inbetween 2 white lines)
- -calibrate O2 monitor (make sure low O2 alarm is working,
- test RA and > 90%)
-check that breathing circuit is complete and undamaged
- -verify CO2 absorbent is adequate (at least ¾ white, a
- little purple is ok)
- -perform breathing system leak check (close APL valve and
- occlude Y piece, pressurize breathing system to 30 cm H20 with O2 flush, ensure
- pressure remains fixed for at least 10 secs, then open APL valve and ensure
- pressure decreases)
- -test vent (place a second breathing bag on Y piece and
- switch to auto (vent) mode, make sure bellows deliver appropriate tidal volume
- and that bag inflates, watch bellows on expiration and ensure bag deflates)
-ensure unidirectional flow valves are working
- -then switch back to manual (bag mode) and ventilate
- manually- assure inflation and deflation and feel appropriate resistance and
- compliance
-check monitors (see above)
- -check final status of machine (vaporizers off, APL valve
- open, selector switch to “bag,” flowmeters at minimum, suction ready, breathing
- system ready to use
-
What should our emergency ventilation equipment consist of
AMBU bag
- Oxygen source separate from the
- anesthesia machine and pipeline supply (AKA O2 cylinder with regulator; also
- need means to open cylinder valve)
-
Stressors of our anesthesia practice
-
What is the AANA wellness program and focus
- ·
- Acknowledges role of internal and external
- stressors
- ·
- Raise awareness and educate
- ·
- Alert students and CRNA’s
- ·
- Wellness = “state of complete physical, mental,
- and social well-being”
- o
- Not only absence of illness but awareness and
- understanding
-
Signs and symptoms of drug abuse and peer assistance
- ·
- 13.4 million alcohol problem
- ·
- 3 million abused/dependent on drugs
- ·
- Chemical dependency = Is defined as a substance
- use disorder characterized by an inability or unwillingness to terminate use in
- spite of serious negative consequences
Signs/Symptoms of abuse:
- ·
- Gradual decline in performance
- ·
- Signs out more narcotics that other providers
- ·
- Inappropriate drug choices
- ·
- Difficulty with authority
- ·
- Forgetful, unpredictable, confused
- ·
- Suffers from tremors or “Monday morning shakes”
- ·
- Appears intoxicated at social functions
- ·
- Discovered comatose or dead
Wellness Program Includes:
- ·
- Assess nature and impact
- ·
- Assisting individuals or organizations
-
. Flow meters :
- Another place oxygen goes is to our flow meters flow meters are calibrated
- for specific gases as Flow rate depends on gas viscosity at how low laminar
- flow and density at high turbulent flow.
- To minimize the effect of friction b/w these (gases and tube wall)
- floats are designed to rotate constantly to keep them in the center of the tube
- Some glow meters have 2 glass tubes, one for low, one for high. The two tubes are in series and controlled by
- one nob.
Dual taper design can allow for reading of both high and low flows.
Malfunctions include:
Debri in flow tube
Vertical flow tube misalignment
Sticking or concealment of flow meter at top of tube
- Flow meter sequence a potential cause of hypoxia in the event of a flow
- meter leak, a potentially dangerous arrangement results when nitrous is located
- in the downstream position. The SAFEST
- configuration exists when o2 is located in the downstream position. (nearest
- the vaporizer) ( This was a mid term
- question!)
- Flow rate depends on gas viscosity at how low laminar flow and density at
- high turbulent flow. To minimize the
- effect of friction b/w these (gas and tube wall) flows are designed to rotate
- constantly to keep them in the center of the tube
- Fabious has convention flow valve but Electronic flow sensor and digital
- displays instead of gas flow tubes
Flow tube shows additive gas flow rate
-
what happens to oxygen
Pathways of O2
1 supply pneumatically powered bellows ventilator
- 2 Via a regulator and an auxiliary O2 flow meter to be connected to a nasal
- cannula, ambu bag, etc
3. To the O2 low pressure alarm sensor
4. Pressure sensitive shutoff valve (fail safe) valve
5. To the O2 flush control valve
6. O2 flow meters
-
Cylinder pressures, pipeline pressures
Cylinder
H Air
Cylinder
Air
O2
N2O
E
1900 PSIG (625L)
1900 PSIG ( 660L)
745 PSIG ( 1590 L)
H
2200 PSIG (6550L)
2200 PSIG (6900L)
745 PSIG (15800L)
E
Pipeline- 50-55 PSIG
-
High pressure systems
High-pressure systems:
- 3.
- Cylinder primary pressure regulator
- 4.
- Cylinder pressure gauge
-
Intermediate pressure system:
- 1.) Cylinder
- supply source (45 psig)
- 3.) O2
- pressure failure devices
-
Low-pressure system:
- 1.) Flow
- meter indicator and tubes
- 3.) Low
- flow pressure reducer regulator (if present)
4.) Vaporizer
-
O2 flush valve
Part of the intermediate pressure circuit
- Ø Allow
- direct communication between O2 high-pressure
- circuit and low-pressure circuit
- Ø Enters
- low-pressure downstream of vaporizers
- Ø Delivers
- 100% O2 at a rate of 35-75L/min
- Ø High-pressure
- O2 source for jet ventilation
Ø Hazards
l Awareness
l barotrauma
-
. what is the PIN index safety system
- Index
- pins- safety (2&5 vs 3&5) pins are on the yolk, holes are on the tank
The Pin Index Safety System- used to prevent mix up of gases
3 & 5: nitrogen oxide
- Issue:
- multiple washers will make the pins obsolete, and you can still have an error
-
O2 nitrous and other failsafe devices
Fail safe valve
- These prevent the delivery of hypoxic (<21% O2) gas
- concentrations.
Fail safe device
- ·
- Permits the flow of other gases (N20, air and
- volatile) only if there is sufficient oxygen pressure to prevent the
- administration of a hypoxic mixture of gases.
-
What is our low pressure alarm
- a.
- Low oxygen pressure alarm
-
- i. Detects
- oxygen supply failure at the common gas inlet
- b.
- Low vent pressure alarm
- 1.
- minimum airway pressure, low airway pressure,
- ventilation failure, apnea, cycling, pressure failure, disconnect, ventilator
- disconnect, minimum ventilatory, ventilation pressure, threshold pressure,
- low-pressure, peak airway, fail-to-cycle, low pressure, low circuit pressure
-
- ii. alarm
- is activated if the pressure detected does not exceed a preset minimum within a
- fixed time
-
1.
Sub atmospheric pressure alarm
- a.
- activated by a pressure that falls below
- atmospheric pressure by a predetermined amount. Subatmospheric pressure can be
- generated by a patient attempting to inhale against a collapsed reservoir bag
- or increased resistance; a blocked inspiratory limb (during the ventilator’s
- expiratory phase); a malfunctioning active closed scavenging system; suction
- applied to a nasogastric tube placed in the tracheobronchial tree or to the
- working channel of an endoscope passed into the airway; a sidestream gas
- analyzer; or the refilling of a hanging bellows ventilator bellows
-
1.
What to do if loss of oxygen pressure
- a.
- Switch over to cylinder and if that doesn’t work
- start to bag patient and obtain another cylinder for O2
-
1.
What is a high pressure alarm
- a.
- Even in the presence of complete obstruction,
- this alarm will not be activated if the peak inspiratory pressure does not
- reach the set limit. High compliance, low resistance, leaks, low inspiratory
- flow rates, high respiratory rates, low I:E ratios, low tidal volumes, and low
- fresh gas flows can all decrease the peak inspiratory pressure so that there is
- no alarm condition. During pressure control ventilation, the inspiratory airway
- pressure is preset and thus cannot act as a warning of tracheal tube occlusion.
|
|